2.Quick detection (only takes time of 2-6s), no dilution needed
3.Auto-detection function is available
4.No need to preheat instrument, measurements can be done upon turning on
5.Long life's Xenon flash lamp,109 flashes
6.Good repeatability due to accurate pathlength control
7.Heating up to 37℃ , stirring speed 100-900rpm for cuevette measurements
8.Only 0.5 ~ 2μl sample volume requested, which can be recycled after test, ideal for precious samples
9.Fluorometer measurement, strong specificity, high sensitivity can make it detect low concentration samples like dsDNA、Oligo、RNA、 protein, down to 0.5pg/μl(dsDNA High Sensitivity)
10.Built-in printer
11.Measurement results can be exported to U disc in format of Excel、JPG
Sample Volume
0.5μl~2μl, advice 2μl
Wavelength Range
180~910nm
Wavelength Accuracy
±1nm
Spectral Resolution
≤ 1.5nm(FWHM@Hg 253.7nm)
Absorbance Range
0.04~750(260nm 波长下,等同于 10mm光程时)
Absorbance Precision
0.002Abs (1mm WaveLength)
Absorbance Accuracy
±1%(7.332 at260nm)
Measurement Time
<6s
Path Length
0.02mm、0.03mm、0.1mm、0.2mm、1mm
Light source / Life
Xenon / Flashes >109
Concentration Range
2 ~37500ng/ul dsDNA;1.32~24750 ng/μL ssDNA ;1.6~30000ng/μL RNA
Protein Detection Range
BSA 0.06 ~ 1119mg/ml;IgG 0.03 ~ 547mg/ml ;A280 0.04~750mg/m1 ;Lysozyme 0.015~284mg/ml
Detector
2048 -element linear CMOS array
Sample pedesta
304 Stainless steel and Quartz fiber
Power Adaptor
DC12V 4A
Power
48W
Dimension(W×D×H)
300x232x196mm
Weight
3.8 kg
OD600 Absorbance Range
0~4.000 Abs
OD600 Absorbance Stability
(0,3)≤ 0.5%, [3,4)≤ 1.5%
OD600 Absorbance Repeatability
(0,3)≤ 0.5%, [3,4)≤ 1.5%
OD600 Absorbance Accuracy
(0,2)≤ 0.005A,[2,3)≤ 1%,[3,4)≤ 2%
OD600 Heating up
37℃
OD600 Stirring speed
100-900rpm
Fluorometer Sample Volume
1-20μl
Fluorometer Measurement Time
3s
Fluorometer Repeatability
< 1.5%
Fluorometer orders of magnitude of concentration
5
Fluorometer Concentration Range
0.01ng/μl-120ng/μl dsDNA HS 0.2ng/μl-2000ng/μl dsDNA BR 0.05ng/μl-240ng/μl Oligo 0.1mg/ml-20mg/ml proteinBR
Fluorometer Linear dynamic range
R2≥ 0.995
Fluorometer Light Source
LED
Fluorometer Excitation wavelength
470/625 (Standard) 365/525 (Optional)
Fluorometer Emission wavelength
525/690 (Standard) 460/620 (Optional)